System Use and Description

The EOI series of primary standard blackbodies are pure elemental freezing point blackbody radiation sources for use as primary standards of spectral radiance. The ITS90 temperature scale is defined by the freezing points of pure elements. EOI freeze point blackbodies provide known spectral radiance at the ITS90 defined temperatures for gallium, indium, tin, zinc, aluminum, silver, gold and copper.

Configuration

The Multi-Point Primary Standard Blackbody system features interchangeable freeze point crucibles which minimizes the investment required to cover the entire ITS90 range of freeze point temps.

The system consists of an award winning EOI temperature controller, crucible holder and one or more of the interchangeable crucibles. The temperature controller provides automatic sequencing of melt and freeze, and a temperature display with freeze indicator. The crucible holder allows the use of interchangeable freeze point crucibles and provides for an optional gas curtain. The crucibles are easily interchanged, allowing testing at multiple freeze point temperatures with one economical system.

Freeze Point Cycle

Freezing Point

Emissivity

Each freeze point blackbody has a 0.25" diameter reverse cone cavity. The reverse cone configuration provides a highest effective emissivity, especially when compared to the more common straight cavity freezing. The cavity configuration and graphite material yield a 0.9997 ±0.0003 effective emissivity.

The graphite reverse conical cavity is surrounded by the pure melt material and takes on the temperature of the metal as it transitions from the liquid to frozen state. This transition or plateau remains constant at ±0.02°C or less. Freeze times are 20 to 40 minutes depending on the melt material, providing the time required to perform testing and calibration of the most sensitive infrared sensors, thermometers, imaging systems and radiometric devices.

EOI primary standard infrared sources have set the performance standard in the most demanding laboratories worldwide since 1964.

Controller

Each MPPS Series blackbody is supplied with a digital 19" rack mount controller.   In the variable temperature mode, the temperature of the blackbody can be set with 0.1°C resolution and actual temperature is displayed to 0.01°C on the front panel of the controller.

LS1050Control.gif (26135 bytes)

Menu Selections Include:

  • Temperature Setting
  • Display Resolution
  • Remote/Local Control
  • K, °C, or °F
  • Ready Window Tolerance
  • Interface Parameters
  • Power On Configuration
  • Self Test
  • Calibration -- Accessed by User Assigned Password

Computer Interfaces

Complete computerized control of the system is available through built-in interfaces.

  • IEEE 488
  • RS-232
  • RS-422
  • RS-485 

All four interfaces are resident in controller. The IEEE 488 and any one of the RS formats may be used at the same time.

Calibration

Each MPPS blackbody is calibrated over the variable operating range at EOI prior to shipment. Recommended calibration cycle is 6 months. Calibration is traceable to NIST.. The calibration table is stored in memory and is easily updated via the front panel keypad.

Case Temperature

The case temperature will never exceed 10°C above ambient.

Operating Ambient

The MPPS blackbody is designed to operate over an ambient temperature range of –40 to +60°C. The MPPS temperature controller will operate over an ambient temperature range of 0 to 50°C.

Options Ordering Suffix
Gas Curtain /GC
Separate Melt Crucibles

Indium

/C150

Tin

/C230

Zinc

/C420

Aluminum

/C660

Silver

/C960

Gold

/C1065

Copper

/C1085

Note: A gas curtain is recommended when operating a graphite cavity above 500 °C.

Standard Models

Model Melt Material Freeze Temp*
(°C)

Melt Material Purity (%)

Variable Temperature 
Range (°C)

MPPS30 Gallium

29.7646**

99.99999

0 - 29

MPPS150 Indium

156.5985

99.9999

50 - 150

MPPS230 Tin

231.928

99.9999

50 - 225

MPPS420 Zinc

419.527

99.9999

50 - 400

MPPS660 Aluminum

660.323

99.99999

50 - 650

MPPS960 Silver

961.78

99.99999

50 - 940

MPPS1065 Gold

1064.18

99.99999

50 - 1000

MPPS1085 Copper

1084.62

99.9999

50 - 1050

* Freezing Temperature Defined by ITS90

**ITS90 defines the gallium temperature as the melting point. The gallium primary standard blackbody utilizes different construction techniques and is a standalone system.

Uncertainty

The uncertainty of the spectral radiance is a function of the emissivity of the conical cavity (geometry plus surface characteristics), wavelength, thermal gradient across the cavity wall, purity of the melt material, uniformity and stability. The calculated uncertainty of the spectral radiance of the EOI freeze point blackbodies is less than 0.1°C (< 0.01°C for Gallium).

Power Requirements

The MPPS blackbodies operate from 100, 120, 220, or 240 VAC ± 10% (switch selectable), 50/60 Hz. The larger systems  require 220V and may require a power amplifier in addition to the controller. Standard cable lengths between controller and blackbody or controller/output stage and output stage/blackbody are 8 feet. The plug-in mains power cable is also 8 feet long.

How to Order 

Select desired model and specify option by adding suffix;

Example:

Model MPPS960 /GC -- Silver Freezing Point with optional gas curtain.