System Use and Description
The EOI LDS series of
differential temperature sources create precise, adjustable thermal
contrast patterns primarily in testing and characterization of IR imaging
systems.
The LDS Series features set point resolution of 0.001°C
and stability of ±0.001°C. These exceptional capabilities enable
accurate evaluation of the most sensitive IR systems.
Configuration
These sources are available in stand alone configuration
or integrated into one of the many Target Projectors/Simulators and
Automatic Test Stations EOI offers. Source sizes (target area with 15°FOV)
range from 1.8" x 1.8" to 12" x 12".
The optional MicroCavity® emitting surface provides a
significant improvement in effective emissivity.
|

|
Controller
Each Laboratory Differential Source is supplied with a
19" rack mount digital controller. Operated manually from the front
panel keypad or automatically using any of the four built in computer
interfaces, this versatile controller also controls all optional
equipment.

Menu Selections Include:
- Temperature Setting
- Control Mode Set
- Differential Mode
- Absolute
- Differential Mode
- Effective Differential
- Display Resolution
- K, °C, °F
- Remote/Local Control
- Ready Window Tolerance
- Interface Parameters
- Power On Configuration
- Self Test
- Calibration:
- Access by User Assigned Password
|
Differential Temperature Mode
Temperature control of the thermal
contrast between controlled surface (source) and target.
Absolute Temperature Mode
Control of source at absolute temperatures.
Effective Differential Mode
Compensation for emissivity differences or losses in adjunct optical
systems so that zero differential temperature corresponds to zero
radiometric contrast.
Computer Interfaces
Complete computerized control of the system is available through
built-in interfaces.
- IEEE 488
- RS-232
- RS-422
- RS-485
All four interfaces are resident in controller. The IEEE 488 and any
one of the RS formats may be used at the same time.
Calibration
Each LDS100 System is calibrated at EOI prior to shipment.
Recommended calibration cycle is 6 months. Calibration is traceable to .
Calibration tables are stored in memory and are easily updated via the
front panel keypad.
Targets
An extensive list of standard four bar, slit and circular targets is
available. EOI also supplies custom targets. Custom targets can be any
pattern with minimum dimension of 0.001" (0.025mm). |